FOR ALL THERMAL NEEDLES

Sensor Body Epoxy
Sensor Needle Stainless Steel
Cable Tefzel, 4 conductor braided copper
Shipping Weight 40 g
Cable Length 1 meter
 


 
KS-1
6 cm long, 1.27 mm diameter
< Best matches theoretical assumptions of the infinite line heat source theory.
Thin enough to accurately measure thermal resistivity without calibration in the field or laboratory.

 
TR-1
10 cm long, 2.4 mm diameter

 

< Conforms to IEEE lab probe length specifications. Factory calibrated for accuracy. If uncalibrated, any thermal needle of this size will give thermal resistivity values about 30% lower than true rho.


 
SH-1
Two needles; 3 cm long, 1.28 mm diameter
 
<
The dual needle probe is used to measure specific heat and diffusivity.
SPECIFICATIONS

KD2 PRO
THERMAL PROPERTIES ANALYZER


Accuracy
5% Thermal Resistivity
5% Specific Heat
10% Thermal Diffusivity

Measurement Speed
1, 2, 5, and 10 minute read times

Compliance to Standard
IEEE 442-1981
ASTM D 5334-08

Range of Measurement

Thermal Resistivity (Rho)
50 to 5000C-cm/W
(SI units: 0.5 to 50mC W-1)
 
Thermal Conductivity (K)
   for KS-1 and SH-1

   0.02 to 2 Wm-1 C-1

   for TR-1
   0.02 to 4 Wm-1 C-1
 
Thermal Diffusivity (D)
0.1 to 1/0 mm2s-1
 
Specific Heat (C)
0.5 to 4 MJ m-3 C-1

Calibration
Each KD2 Pro comes factory calibrated and includes performance verification
standards.




Type
Ultra low-power 16-bit microcontroller with 24-bit A/D converter

Case Size
15.5 x 9.5 x 3.5 cm

Battery Source
4 AA

Screen Type
7.5 cm x 4 cm LCD

Auto-Read Mode
Long-term monitoring option allows unattended data collection at user-defined intervals.

Data Storage
4095 readings; flash memory

Included Accessories
User's manual, Pelican carrying case, readout stand, performance verification standards, thermal grease, custom drill bit for drilling pilot holes in rock and concrete samples, KD2 Pro download utility, RS232 cable.